A simple, precise and accurate reverse phase high performance liquid chromatographic method was developed for the simultaneous estimation of guaiphenesin and phenylephrine in their tablet dosage form. The chromatographic conditions were standardizedusing anInertsilcolumn,C18 (150x4.6 ID) 5μm with UV detection at 277nm , and the mobile phase consisted of sodium dihydrogenphosphate buffer: Acetonitrile (30:70).The retention times of guaiphenesin and phenylephrine were About 3.497 min for Guaifenesin and 2.353min for Phenylephrine, respectively. The calibration curves were linear with correlation coefficients of 0.9987, 0.9988, 0.9981 and 0.9981 over a concentration range of 4.0–24.0 μg/ml for guaiphenesin, 5.0–30.0 μg/ml for phenylephrine, respectively. The proposed method has been validated according to the ICH guidelines and was successfully applied to estimate the levels of two drugs in a combined formulation with good accuracy and precision